SC-5A Electrical Ventilator |
| Main Specifications ,Ventilation modes: IPPV, IPPV/Assist ,Ventilation frequency: 6--60 cycles/seconds ,I:E ratio: 1:1,1:1.5,1:2,1:3 ,Tidal volume: 200~1200/ml ,Oxygen concentration: 21%--100% ,High Pressure alarm: 2--6kPa ,Low Pressure alarm: 0.4--2.4kPa ,Trigger sensitivity: -0.2KPa ,PEEP: 0-20 cmH2O(Optional) ,Maximum safety pressure: 6.0KPa ,Environmental requirement: Normal working condition Surrounding temperature:5--40C Surrounding humidity: <80% ,Aero-pressure: 86-106KPa ,Power supply AC 220V 50Hz Monitor Function | |
|
SC-5A Electrical Ventilator is a time electrical-driven, electronic-controlled, time cycled and volume constant ventilator for the respiration administration and theaply to the patient with ARDS or respiratory failure. This ventilator configures two ventilation modes of IPPV and IPPV/ASSIST, and monitoring functions as high and low airway pressure, wrong sestting and etc. The alarm is visible and audible. The ventilator is not only a state of multi-functions but also a great convenience in manipulation. Because of electrical driven and electronic-controlledl, the ventilator needs no compressed air, and the work is at low noice. | |