Ultrasound Scanner -- 3000E1 B

This B ultrasound system is a portable electronic convex scanning ultrasound tomography inspection device with digital scanning conversion. It has the features of small bulk, light weight, large LCD screen display and no eradiation.

Features:            

  • 15'' LCD (abbr. Liquid Crystal Display)
  • Multi-frequency probes
  • Real-time zooming
  • 128-frame cineloop
  • Multi-segment TGC adjustment

 

Imaging Processing:            

  • Part zoom in real time, gray curve control, STC curve, edge enhancement, dynamic range, frame correlation, 256 gray scale, image reverse function
  • Pre-processing: multi-segments TGC adjustment, frame correlation, image direction
  • Post-processing: left/right, up/down, positive/negative
  • Zoom: magnification (4 steps)
  • Cine loop: 128 frames

 

Software Functions:            

  • General measurement: distance, circumference, area, oval, trace, volume, HR, gestational week
  • Obstetric measurement soft: GS, CRL, BPD, AC, HC, FL, AFI, EDD, and calculating fetus weight, pregnancy week, expected delivery date, fetal physiological rating according to the previous measurements

 

Peripheral Ports:             

  • VGA port
  • PAL-D video output

 

Standard Configuration:             

  • Main unit
  • 15'' LCD (abbr. Liquid Crystal Display)
  • 128-frame cine loop
  • one probe connectors
  • 3.5MHz electronic convex probe (2.5/5.0MHz)

 

Optional:             

  • Electronic transvaginal probe (5.0/7.5MHz)
  • High-frequency linear probe (6.5/8.5MHz)
  • Video printer

 

Specification:

Image mode

B, B/B, B+M, M, 4B, 8B

Scanning mode

Convex and linear array

Multi-frequency probe

2.5~8.5 MHz

Image magnification

¡Á1.0, ¡Á1.2, ¡Á1.5, ¡Á2.0.

Electronic focus

Acoustic lens focus, single focus and multiple focus

Max display depth

220mm

Depth enhancement

Yes

Local zoom

Yes

Frame correlation function

Yes

Adjustment total gain and dynamic range

Yes

Gray Scale

256

Display

15'' LCD

Gain Control

6-segment STC segmentation control,  full gain control

Probe connectors

1 (standard)

Cine Loop

128 frames

Scanning frequency

2.5~8.5 MHz (depends on probes)

Image reverse

Left/right, up/down, positive/negative

Comment

Name, Age, Sex£¬number of patient recorder, diagnostic date, 24 markers of body location, remarks in full screen

Body marks

18

Gray Scale Conversion

¦Ã rectification

Pseudo Color Function

8 colors

Power supply

110-240V AC, 50Hz¡À1Hz

Power consumption

¡Ü100VA

Dimensions

380x285x110mm (LxHxM)

Net weight

approx. 6.8kg

¡¡